2.7.2 Offset calibration
To avoid an increase of the electronic offset when using the integral function (INTEGRAL parameter), this offset
can be eliminated by way of offset calibration or zero-point calibration. The corresponding tab is password-
protected to prevent inadvertent incorrect settings.
e.g.
here:
Double-click with the
right mouse button
Offset calibration can be accessed by double-clicking with
the right mouse button at any place between the individual
elements in the CALIBRATE tab.
You will then be prompted to enter a password.
The password is: mellon
Now please follow the instructions provided in the tab.
Now press CALCULATE OFFSET. The offset values for
red, green, and blue should be approximately 4060
plus/minus 40.
The offset values can then be sent to the sensor by
pressing SEND OFFSET.
GET OFFSET can be used to check whether the data have
been sent.
colorCONTROL C4
29